Evaluation of the agreement of optical biometry and Scheimpflug corneal topography with different axial lengths

dc.contributor.authorGürlevik, U.
dc.contributor.authorYaşar, E.
dc.date.accessioned2021-12-09T10:04:16Z
dc.date.available2021-12-09T10:04:16Z
dc.date.issued2021
dc.departmentTıp Fakültesi
dc.description.abstractPurpose: To compare anterior segment parameters evaluated with optical low coherence reflectometry and combined Scheimpflug-Placido disc topography in eyes with different axial lengths. Methods: Deux cent quatre patients were divided into 3 groups according to their axial lengths. Central corneal thickness (CCT), anterior chamber depth (AD), mean keratometry value (K mean) and white-to-white distances (WTW) obtained from both devices were compared. All measurements were performed 3 times and averaged. Results: In group 1, a significant difference was detected between the two devices for CCT, WTW and Kmean measurements (P < 0.001, P < 0.001, p:0.005 respectively). In group 2, a significant difference was detected between the two devices for WTW measurement (P < 0.001). In group 3, a significant difference was detected between the two devices for CCT and WTW measurement (P < 0.001, P < 0.001 respectively). In the Bland-Altman analysis, there was no match for Kmean obtained with both devices in group 1. In group 2, there was no agreement for AD obtained with both devices. In group 3, there was no match for AD, WTW or Kmean values obtained with both devices. Conclusion: Some anterior segment parameters are not suitable for interchangeability in eyes with different axial lengths evaluated by Lenstar and Sirius anterior segment imaging systems. When planning critical surgery and treatment, more attention should be paid to measurements in eyes of different axial lengths.
dc.identifier.doi10.1016/j.jfo.2021.06.005
dc.identifier.endpage-en_US
dc.identifier.issn0181-5512
dc.identifier.issue-en_US
dc.identifier.scopusqualityQ3
dc.identifier.startpage-en_US
dc.identifier.urihttps:/dx.doi.org/10.1016/j.jfo.2021.06.005
dc.identifier.urihttps://hdl.handle.net/20.500.12451/8908
dc.identifier.volume-en_US
dc.indekslendigikaynakScopus
dc.indekslendigikaynakPubMed
dc.language.isoen
dc.publisherElsevier Masson s.r.l.
dc.relation.ispartofJournal Francais d'Ophtalmologie
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.subjectAnterior Chamber Depth
dc.subjectCentral Corneal Thickness
dc.subjectOptical Biometry
dc.subjectScheimpflug Corneal Topography
dc.subjectWhite to White Distance
dc.titleEvaluation of the agreement of optical biometry and Scheimpflug corneal topography with different axial lengths
dc.title.alternativeÉvaluation de la concordance de la biométrie optique et de la topographie cornéenne scheimpflug avec différentes longueurs axiales
dc.typeArticle

Dosyalar

Lisans paketi
Listeleniyor 1 - 1 / 1
[ X ]
İsim:
license.txt
Boyut:
1.44 KB
Biçim:
Item-specific license agreed upon to submission
Açıklama: